BE OVP (over-voltage protection) cycling test.
|
|
|
|
Coordinates four Modbus RTU devices, each on its own COM port:
|
|
|
|
RCU - Beyond receiver unit (a.k.a. RPU). Charging command + telemetry.
|
|
TPU - Beyond transmitter unit. Telemetry (mirrors RCU values over RF).
|
|
RELAY - Phoenix/Fenghuang-family board carrying the PFC relay used to
|
|
create the over-voltage condition (coil 12 = PFC_relay_enable).
|
|
JIG - XYZ positioner.
|
|
|
|
The RCU is powered THROUGH the relay: with the relay open it is dead and
|
|
answers nothing. So the relay is closed and given RELAY_SETTLE_S to boot the
|
|
RCU before any RCU access anywhere in this script - the link test included -
|
|
and while the relay is open the RCU telemetry columns are left blank.
|
|
|
|
Per cycle:
|
|
1. Relay control = 1, then wait RELAY_SETTLE_S for the RCU to power up.
|
|
2. Jig -> NEAR_POSITION, settle.
|
|
3. RCU charging cmd = 1, then wait for RCU main_state == CHARGING(5) and
|
|
output current above the threshold held continuously for the dwell.
|
|
Abort the cycle on timeout.
|
|
4. Relay control = 0 (this also cuts RCU power).
|
|
5. Jig -> FAR_POSITION, keep monitoring for FAR_OBSERVE_S.
|
|
6. RCU charging cmd = 0 (best-effort; the RCU is unpowered by now),
|
|
settle, next cycle.
|
|
|
|
A background monitor thread samples RCU and TPU input registers every 5 s for
|
|
the whole run (all cycles, including the aborted ones) and appends each sample
|
|
to the CSV.
|
|
|
|
Usage:
|
|
# check the wiring: read registers from all four ports, then exit
|
|
python BE-RCU-ovp_test_csv.py -r COM4 -t COM10 -l COM13 -j COM9 --link-test
|
|
|
|
# run the test (the link test also runs as a pre-flight check)
|
|
python BE-RCU-ovp_test_csv.py -r COM4 -t COM10 -l COM13 -j COM9 -n 1
|
|
"""
|
|
|
|
import argparse
|
|
import csv
|
|
import datetime
|
|
import logging
|
|
import os
|
|
import sys
|
|
import threading
|
|
import time
|
|
|
|
import minimalmodbus
|
|
|
|
|
|
# ---------------------------------------------------------------------------
|
|
# Serial / Modbus link
|
|
# ---------------------------------------------------------------------------
|
|
SLAVE_ID = 0x10 # all four devices share slave id 16
|
|
BAUDRATE = 9600
|
|
SERIAL_TIMEOUT = 2.0 # seconds
|
|
|
|
|
|
# ---------------------------------------------------------------------------
|
|
# Register map
|
|
#
|
|
# Sources (in order of authority):
|
|
# [1] be-rssi-auto-en/charging_test.py - RCU/TPU "beyond" family classes
|
|
# [2] be-rssi-auto-en/base-definition.md - BE system register spec
|
|
# [3] be-rssi-auto-en/be_corner_test.py - working BE jig/charging test
|
|
# [4] pegasus/config-fenghuang/config.json - phoenix-family board (relay)
|
|
# [5] df-code/main.cpp - firmware-side main_state enum
|
|
#
|
|
# All telemetry below is read with function code 4 (input registers); commands
|
|
# are coils (function code 5) and jig positions are holding registers (fc 6).
|
|
# ---------------------------------------------------------------------------
|
|
|
|
# --- RCU / RPU, "beyond" family [1][2] -------------------------------------
|
|
RCU_CHARGING_CMD_ADDRESS = 0 # coil - 1 = start charging, 0 = stop
|
|
RCU_RANGE_CHECK_CMD_ADDRESS = 2 # coil - not used by this test
|
|
RCU_SHADOW_ERROR_LO_ADDRESS = 0 # input
|
|
RCU_SHADOW_ERROR_HI_ADDRESS = 1 # input
|
|
RCU_DC_BUS_VOLTAGE_ADDRESS = 30 # input, Q128 (128 counts = 1 V)
|
|
RCU_OUTPUT_CURRENT_ADDRESS = 31 # input, Q128 (128 counts = 1 A)
|
|
RCU_MAIN_PRE_STATE_ADDRESS = 50 # input
|
|
RCU_MAIN_STATE_ADDRESS = 51 # input
|
|
RCU_ERROR_CODE_LO_ADDRESS = 52 # input
|
|
RCU_ERROR_CODE_HI_ADDRESS = 53 # input
|
|
RCU_OUTPUT_VOLTAGE_ADDRESS = 106 # input, Q128 (phoenix family uses 32)
|
|
Q128_DIVISOR = 128.0
|
|
|
|
# --- TPU, "beyond" family [1][2] -------------------------------------------
|
|
TPU_PRI_ENABLE_RF_ADDRESS = 30 # coil - not used by this test
|
|
TPU_DC_BUS_VOLTAGE_ADDRESS = 38 # input, float32 (2 regs, big-endian)
|
|
TPU_OUTPUT_CURRENT_ADDRESS = 40 # input, float32
|
|
TPU_MAIN_PRE_STATE_ADDRESS = 49 # input
|
|
TPU_MAIN_STATE_ADDRESS = 50 # input
|
|
TPU_ERROR_CODE_LO_ADDRESS = 61 # input
|
|
TPU_ERROR_CODE_HI_ADDRESS = 62 # input
|
|
TPU_SHADOW_ERROR_LO_ADDRESS = 63 # input
|
|
TPU_SHADOW_ERROR_HI_ADDRESS = 64 # input
|
|
TPU_OUTPUT_VOLTAGE_ADDRESS = 74 # input, float32
|
|
TPU_RCU_ERROR_CODE_HI_ADDRESS = 134 # input (note: hi is the LOWER address)
|
|
TPU_RCU_ERROR_CODE_LO_ADDRESS = 135 # input
|
|
|
|
# --- RELAY board, phoenix/fenghuang family [4] -----------------------------
|
|
RELAY_CONTROL_CMD_ADDRESS = 12 # coil - PFC_relay_enable, 1 = on
|
|
RELAY_MAIN_PRE_STATE_ADDRESS = 60 # input
|
|
RELAY_MAIN_STATE_ADDRESS = 61 # input
|
|
|
|
# --- JIG [1][2][3] ---------------------------------------------------------
|
|
JIG_CMD_ADDRESS = 30 # coil - 1 = move to position
|
|
JIG_X_ADDRESS = 90 # holding - mm, signed
|
|
JIG_Y_ADDRESS = 91 # holding - mm, signed
|
|
JIG_Z_ADDRESS = 92 # holding - mm, signed
|
|
|
|
# --- RCU main_state enum [5] -----------------------------------------------
|
|
RCU_STATE_IDLE = 1
|
|
RCU_STATE_HANDSHAKE = 2
|
|
RCU_STATE_PAIRED = 3
|
|
RCU_STATE_PRE_CHARGING = 4
|
|
RCU_STATE_CHARGING = 5
|
|
RCU_STATE_STOP = 6
|
|
RCU_STATE_ERROR = 238
|
|
RCU_STATE_CONFIGURATION = 240
|
|
|
|
|
|
# ---------------------------------------------------------------------------
|
|
# Test profile
|
|
# ---------------------------------------------------------------------------
|
|
NEAR_POSITION = (-30, 0, 20) # (x, y, z) mm - charging position
|
|
FAR_POSITION = (30, 0, 20) # (x, y, z) mm - move-away position
|
|
|
|
JIG_SETTLE_S = 5.0 # after the near move, before the relay
|
|
CHARGING_TIMEOUT_S = 60.0 # max wait for the charging criteria
|
|
CHARGING_DWELL_S = 30.0 # current must stay above threshold this long
|
|
CHARGING_CURRENT_THRESHOLD_A = 3.0
|
|
FAR_OBSERVE_S = 15.0 # keep monitoring after the move-away
|
|
CYCLE_GAP_S = 5.0 # after charging cmd = 0
|
|
MONITOR_INTERVAL_S = 2.0 # telemetry sampling / CSV row period
|
|
RELAY_SETTLE_S = 10.0 # time to wait for relay to settle
|
|
|
|
RESULTS_DIR = "results"
|
|
|
|
logger = logging.getLogger("ovp")
|
|
|
|
|
|
# ---------------------------------------------------------------------------
|
|
# Modbus plumbing
|
|
# ---------------------------------------------------------------------------
|
|
class Device:
|
|
"""One Modbus RTU device on its own serial port.
|
|
|
|
Every access takes the per-device lock: the monitor thread reads telemetry
|
|
while the main thread writes commands, and minimalmodbus instruments are
|
|
not safe for concurrent use.
|
|
"""
|
|
|
|
def __init__(self, port, name, slave_id=SLAVE_ID):
|
|
self.name = name
|
|
self.port = port
|
|
self._lock = threading.Lock()
|
|
self.inst = minimalmodbus.Instrument(
|
|
port=port, slaveaddress=slave_id, mode='rtu')
|
|
self.inst.serial.baudrate = BAUDRATE
|
|
self.inst.serial.timeout = SERIAL_TIMEOUT
|
|
self.inst.close_port_after_each_call = False
|
|
logger.info("%s connected on %s (slave id %d)", name, port, slave_id)
|
|
|
|
# Registers read by the link test, as (label, address, kind).
|
|
PROBE_REGISTERS = ()
|
|
|
|
# --- raw accessors -----------------------------------------------------
|
|
def read_u16(self, address):
|
|
with self._lock:
|
|
return self.inst.read_register(address, functioncode=4)
|
|
|
|
def read_holding_u16(self, address):
|
|
with self._lock:
|
|
return self.inst.read_register(address, functioncode=3)
|
|
|
|
def read_f32(self, address):
|
|
"""Read a big-endian IEEE-754 float from two input registers."""
|
|
with self._lock:
|
|
return self.inst.read_float(
|
|
address, functioncode=4, number_of_registers=2)
|
|
|
|
def write_coil(self, address, value):
|
|
with self._lock:
|
|
self.inst.write_bit(address, 1 if value else 0, functioncode=5)
|
|
|
|
def write_holding(self, address, value):
|
|
# Jig coordinates may be negative; Modbus registers are unsigned.
|
|
with self._lock:
|
|
self.inst.write_register(address, value & 0xFFFF, functioncode=6)
|
|
|
|
def probe(self):
|
|
"""Read PROBE_REGISTERS one at a time to prove the link is alive.
|
|
|
|
Returns a list of (label, address, kind, value, error) where exactly
|
|
one of value/error is None.
|
|
"""
|
|
reads = []
|
|
for label, address, kind in self.PROBE_REGISTERS:
|
|
try:
|
|
if kind == 'input':
|
|
value = self.read_u16(address)
|
|
elif kind == 'holding':
|
|
value = self.read_holding_u16(address)
|
|
elif kind == 'holding_signed':
|
|
raw = self.read_holding_u16(address)
|
|
value = raw - 0x10000 if raw > 0x7FFF else raw
|
|
elif kind == 'float':
|
|
value = round(self.read_f32(address), 3)
|
|
else:
|
|
raise ValueError("unknown probe kind: %s" % kind)
|
|
reads.append((label, address, kind, value, None))
|
|
except Exception as exc:
|
|
reads.append((label, address, kind, None, str(exc)))
|
|
return reads
|
|
|
|
def close(self):
|
|
try:
|
|
self.inst.serial.close()
|
|
logger.info("%s disconnected", self.name)
|
|
except Exception as exc:
|
|
logger.warning("%s: close failed: %s", self.name, exc)
|
|
|
|
|
|
def _safe(fn, *args):
|
|
"""Run a read and return None instead of raising, so one bad register
|
|
never kills the monitor thread mid-run."""
|
|
try:
|
|
return fn(*args)
|
|
except Exception as exc:
|
|
logger.debug("read failed: %s", exc)
|
|
return None
|
|
|
|
|
|
def _combine_error(lo, hi):
|
|
"""Join the lo/hi register halves into one 32-bit code as 8 hex digits.
|
|
|
|
Format matches be-rssi-auto-en/charging_test.py and the Beyond error code
|
|
map, so a value can be looked up there directly. An unreadable register
|
|
gives "" rather than 00000000, which would read as "no error".
|
|
"""
|
|
if lo is None or hi is None:
|
|
return ""
|
|
return "%08X" % ((hi << 16) | lo)
|
|
|
|
|
|
def _q128(raw):
|
|
return None if raw is None else round(raw / Q128_DIVISOR, 3)
|
|
|
|
|
|
def _round(value, digits=3):
|
|
return None if value is None else round(value, digits)
|
|
|
|
|
|
# ---------------------------------------------------------------------------
|
|
# Devices
|
|
# ---------------------------------------------------------------------------
|
|
# What a sample looks like when the RCU is unpowered (relay open).
|
|
BLANK_RCU_DATA = {
|
|
'main_pre_state': None,
|
|
'main_state': None,
|
|
'output_current': None,
|
|
'output_voltage': None,
|
|
'dc_bus_voltage': None,
|
|
'error_code': '',
|
|
'shadow_error': '',
|
|
}
|
|
|
|
|
|
class RCU(Device):
|
|
PROBE_REGISTERS = (
|
|
('main_state', RCU_MAIN_STATE_ADDRESS, 'input'),
|
|
('main_pre_state', RCU_MAIN_PRE_STATE_ADDRESS, 'input'),
|
|
('output_current_q128', RCU_OUTPUT_CURRENT_ADDRESS, 'input'),
|
|
('output_voltage_q128', RCU_OUTPUT_VOLTAGE_ADDRESS, 'input'),
|
|
)
|
|
|
|
def __init__(self, port):
|
|
super().__init__(port, "RCU")
|
|
|
|
def set_charging(self, enable):
|
|
logger.info("RCU charging cmd = %d", 1 if enable else 0)
|
|
self.write_coil(RCU_CHARGING_CMD_ADDRESS, enable)
|
|
|
|
def read_data(self):
|
|
return {
|
|
'main_pre_state': _safe(self.read_u16, RCU_MAIN_PRE_STATE_ADDRESS),
|
|
'main_state': _safe(self.read_u16, RCU_MAIN_STATE_ADDRESS),
|
|
'output_current': _q128(
|
|
_safe(self.read_u16, RCU_OUTPUT_CURRENT_ADDRESS)),
|
|
'output_voltage': _q128(
|
|
_safe(self.read_u16, RCU_OUTPUT_VOLTAGE_ADDRESS)),
|
|
'dc_bus_voltage': _q128(
|
|
_safe(self.read_u16, RCU_DC_BUS_VOLTAGE_ADDRESS)),
|
|
'error_code': _combine_error(
|
|
_safe(self.read_u16, RCU_ERROR_CODE_LO_ADDRESS),
|
|
_safe(self.read_u16, RCU_ERROR_CODE_HI_ADDRESS)),
|
|
'shadow_error': _combine_error(
|
|
_safe(self.read_u16, RCU_SHADOW_ERROR_LO_ADDRESS),
|
|
_safe(self.read_u16, RCU_SHADOW_ERROR_HI_ADDRESS)),
|
|
}
|
|
|
|
|
|
class TPU(Device):
|
|
PROBE_REGISTERS = (
|
|
('main_state', TPU_MAIN_STATE_ADDRESS, 'input'),
|
|
('main_pre_state', TPU_MAIN_PRE_STATE_ADDRESS, 'input'),
|
|
('rcu_output_current', TPU_OUTPUT_CURRENT_ADDRESS, 'float'),
|
|
('rcu_output_voltage', TPU_OUTPUT_VOLTAGE_ADDRESS, 'float'),
|
|
)
|
|
|
|
def __init__(self, port):
|
|
super().__init__(port, "TPU")
|
|
|
|
def read_data(self):
|
|
return {
|
|
'main_pre_state': _safe(self.read_u16, TPU_MAIN_PRE_STATE_ADDRESS),
|
|
'main_state': _safe(self.read_u16, TPU_MAIN_STATE_ADDRESS),
|
|
'output_current': _round(
|
|
_safe(self.read_f32, TPU_OUTPUT_CURRENT_ADDRESS)),
|
|
'output_voltage': _round(
|
|
_safe(self.read_f32, TPU_OUTPUT_VOLTAGE_ADDRESS)),
|
|
'dc_bus_voltage': _round(
|
|
_safe(self.read_f32, TPU_DC_BUS_VOLTAGE_ADDRESS)),
|
|
'error_code': _combine_error(
|
|
_safe(self.read_u16, TPU_ERROR_CODE_LO_ADDRESS),
|
|
_safe(self.read_u16, TPU_ERROR_CODE_HI_ADDRESS)),
|
|
'shadow_error': _combine_error(
|
|
_safe(self.read_u16, TPU_SHADOW_ERROR_LO_ADDRESS),
|
|
_safe(self.read_u16, TPU_SHADOW_ERROR_HI_ADDRESS)),
|
|
'rcu_error_code': _combine_error(
|
|
_safe(self.read_u16, TPU_RCU_ERROR_CODE_LO_ADDRESS),
|
|
_safe(self.read_u16, TPU_RCU_ERROR_CODE_HI_ADDRESS)),
|
|
}
|
|
|
|
|
|
class Relay(Device):
|
|
PROBE_REGISTERS = (
|
|
('main_state', RELAY_MAIN_STATE_ADDRESS, 'input'),
|
|
('main_pre_state', RELAY_MAIN_PRE_STATE_ADDRESS, 'input'),
|
|
)
|
|
|
|
def __init__(self, port):
|
|
super().__init__(port, "RELAY")
|
|
|
|
def set_control(self, enable):
|
|
logger.info("RELAY control cmd = %d", 1 if enable else 0)
|
|
self.write_coil(RELAY_CONTROL_CMD_ADDRESS, enable)
|
|
|
|
def read_data(self):
|
|
return {'main_state': _safe(self.read_u16, RELAY_MAIN_STATE_ADDRESS)}
|
|
|
|
|
|
class Jig(Device):
|
|
# The jig exposes no input registers, so the link test reads its position
|
|
# holding registers back instead - harmless, and it proves the link.
|
|
PROBE_REGISTERS = (
|
|
('x_position_mm', JIG_X_ADDRESS, 'holding_signed'),
|
|
('y_position_mm', JIG_Y_ADDRESS, 'holding_signed'),
|
|
('z_position_mm', JIG_Z_ADDRESS, 'holding_signed'),
|
|
)
|
|
|
|
def __init__(self, port):
|
|
super().__init__(port, "JIG")
|
|
|
|
def move_to(self, x, y, z):
|
|
logger.info("JIG move to (%d, %d, %d) mm", x, y, z)
|
|
self.write_holding(JIG_X_ADDRESS, x)
|
|
self.write_holding(JIG_Y_ADDRESS, y)
|
|
self.write_holding(JIG_Z_ADDRESS, z)
|
|
self.write_coil(JIG_CMD_ADDRESS, True)
|
|
|
|
|
|
# ---------------------------------------------------------------------------
|
|
# RCU power
|
|
# ---------------------------------------------------------------------------
|
|
def power_up_rcu(relay, settle_s=None):
|
|
"""Close the relay and wait for the RCU to boot.
|
|
|
|
The RCU is fed through the relay, so it is dead - and will not answer a
|
|
single Modbus request - until the relay is on. Every RCU access, reads
|
|
included, has to come after this.
|
|
"""
|
|
# Read the constant here, not as a default, so it stays overridable.
|
|
settle_s = RELAY_SETTLE_S if settle_s is None else settle_s
|
|
relay.set_control(True)
|
|
logger.info("Relay closed - waiting %.0fs for the RCU to power up...",
|
|
settle_s)
|
|
time.sleep(settle_s)
|
|
|
|
|
|
# ---------------------------------------------------------------------------
|
|
# Link test
|
|
# ---------------------------------------------------------------------------
|
|
def link_test(devices):
|
|
"""Read a handful of registers from every device and report per port.
|
|
|
|
Returns (all_ok, any_dead): all_ok is False if any single register could
|
|
not be read, any_dead is True if some device answered nothing at all -
|
|
which usually means a wrong or swapped COM port.
|
|
"""
|
|
logger.info("-" * 62)
|
|
logger.info("LINK TEST")
|
|
|
|
all_ok = True
|
|
any_dead = False
|
|
|
|
for dev in devices:
|
|
reads = dev.probe()
|
|
good = sum(1 for r in reads if r[4] is None)
|
|
if good == len(reads):
|
|
status = "OK"
|
|
elif good:
|
|
status = "PARTIAL"
|
|
all_ok = False
|
|
else:
|
|
status = "NO REPLY"
|
|
all_ok = False
|
|
any_dead = True
|
|
|
|
logger.info(" %-6s %-10s %-9s %d/%d registers",
|
|
dev.name, dev.port, status, good, len(reads))
|
|
for label, address, kind, value, err in reads:
|
|
if err is None:
|
|
logger.info(" %-21s %-14s @%-4d = %s",
|
|
label, kind, address, value)
|
|
else:
|
|
logger.warning(" %-21s %-14s @%-4d ! %s",
|
|
label, kind, address, err)
|
|
|
|
logger.info("-" * 62)
|
|
return all_ok, any_dead
|
|
|
|
|
|
def run_link_test(rcu_port, tpu_port, relay_port, jig_port):
|
|
"""Stand-alone link test: open every port, probe it, close it.
|
|
|
|
Returns True only if all four devices answered every probe.
|
|
"""
|
|
devices = []
|
|
connect_ok = True
|
|
|
|
wanted = [(RCU, rcu_port), (TPU, tpu_port), (Relay, relay_port)]
|
|
if jig_port:
|
|
wanted.append((Jig, jig_port))
|
|
else:
|
|
logger.info("No jig port given - skipping the jig.")
|
|
|
|
for cls, port in wanted:
|
|
try:
|
|
devices.append(cls(port))
|
|
except Exception as exc:
|
|
logger.error("%s on %s: connect failed: %s",
|
|
cls.__name__.upper(), port, exc)
|
|
connect_ok = False
|
|
|
|
relay = next((d for d in devices if isinstance(d, Relay)), None)
|
|
|
|
try:
|
|
# The RCU has no power until the relay is on, so it could not answer
|
|
# a probe before this.
|
|
if relay is not None:
|
|
power_up_rcu(relay)
|
|
else:
|
|
logger.warning("No relay - the RCU is unpowered, so its probe "
|
|
"will fail regardless of wiring.")
|
|
|
|
all_ok, _any_dead = link_test(devices) if devices else (False, True)
|
|
finally:
|
|
# Leave the rig de-energised again.
|
|
if relay is not None:
|
|
try:
|
|
relay.set_control(False)
|
|
except Exception as exc:
|
|
logger.warning("Could not switch relay off: %s", exc)
|
|
for dev in devices:
|
|
dev.close()
|
|
|
|
result = connect_ok and all_ok
|
|
logger.info("Link test: %s", "PASS" if result else "FAIL")
|
|
return result
|
|
|
|
|
|
# ---------------------------------------------------------------------------
|
|
# CSV
|
|
# ---------------------------------------------------------------------------
|
|
class CSVRecorder:
|
|
HEADERS = [
|
|
'Timestamp', 'Elapsed_s', 'Cycle', 'Phase',
|
|
'Jig_X_mm', 'Jig_Y_mm', 'Jig_Z_mm',
|
|
'Relay_Cmd', 'RCU_Charging_Cmd', 'RCU_Powered',
|
|
'RCU_Main_Pre_State', 'RCU_Main_State',
|
|
'RCU_Output_Current_A', 'RCU_Output_Voltage_V', 'RCU_DC_Bus_Voltage_V',
|
|
'RCU_Error_Code', 'RCU_Shadow_Error',
|
|
'TPU_Main_Pre_State', 'TPU_Main_State',
|
|
'TPU_Output_Current_A', 'TPU_Output_Voltage_V', 'TPU_DC_Bus_Voltage_V',
|
|
'TPU_Error_Code', 'TPU_Shadow_Error', 'TPU_RCU_Error_Code',
|
|
'Relay_Main_State', 'Charging_Dwell_s', 'Note',
|
|
]
|
|
|
|
def __init__(self, path):
|
|
self.path = path
|
|
self._lock = threading.Lock()
|
|
self._file = open(path, 'w', newline='', encoding='utf-8')
|
|
self._writer = csv.writer(self._file)
|
|
self._writer.writerow(self.HEADERS)
|
|
self._file.flush()
|
|
|
|
def write(self, row):
|
|
with self._lock:
|
|
self._writer.writerow(row)
|
|
self._file.flush()
|
|
|
|
def close(self):
|
|
with self._lock:
|
|
self._file.close()
|
|
|
|
|
|
# ---------------------------------------------------------------------------
|
|
# Monitor thread
|
|
# ---------------------------------------------------------------------------
|
|
class Monitor(threading.Thread):
|
|
"""Samples RCU/TPU/relay telemetry on a fixed interval for the whole run.
|
|
|
|
It also tracks how long the charging criteria have held continuously, so
|
|
the main thread never has to touch the serial ports while it waits.
|
|
"""
|
|
|
|
def __init__(self, rcu, tpu, relay, recorder, t0,
|
|
interval=MONITOR_INTERVAL_S,
|
|
current_threshold=CHARGING_CURRENT_THRESHOLD_A):
|
|
super().__init__(name="monitor", daemon=True)
|
|
self.rcu = rcu
|
|
self.tpu = tpu
|
|
self.relay = relay
|
|
self.recorder = recorder
|
|
self.t0 = t0
|
|
self.interval = interval
|
|
self.current_threshold = current_threshold
|
|
|
|
self._stop = threading.Event()
|
|
self._state_lock = threading.Lock()
|
|
self._cycle = 0
|
|
self._phase = "startup"
|
|
self._note = ""
|
|
self._jig = (None, None, None)
|
|
self._relay_cmd = 0
|
|
self._charging_cmd = 0
|
|
self._rcu_powered = False
|
|
self._ok_since = None
|
|
self._latest = {}
|
|
|
|
# --- context set by the main thread ------------------------------------
|
|
def set_context(self, cycle=None, phase=None, jig=None,
|
|
relay_cmd=None, charging_cmd=None, note=None,
|
|
rcu_powered=None):
|
|
with self._state_lock:
|
|
if cycle is not None:
|
|
self._cycle = cycle
|
|
if phase is not None:
|
|
self._phase = phase
|
|
if jig is not None:
|
|
self._jig = tuple(jig)
|
|
if relay_cmd is not None:
|
|
self._relay_cmd = int(bool(relay_cmd))
|
|
if charging_cmd is not None:
|
|
self._charging_cmd = int(bool(charging_cmd))
|
|
if note is not None:
|
|
self._note = note
|
|
if rcu_powered is not None:
|
|
self._rcu_powered = bool(rcu_powered)
|
|
|
|
def reset_dwell(self):
|
|
with self._state_lock:
|
|
self._ok_since = None
|
|
|
|
def dwell_seconds(self):
|
|
"""Seconds the charging criteria have held continuously (0 if not)."""
|
|
with self._state_lock:
|
|
if self._ok_since is None:
|
|
return 0.0
|
|
return time.time() - self._ok_since
|
|
|
|
def latest(self):
|
|
with self._state_lock:
|
|
return dict(self._latest)
|
|
|
|
def stop(self):
|
|
self._stop.set()
|
|
|
|
# --- loop --------------------------------------------------------------
|
|
def run(self):
|
|
while not self._stop.is_set():
|
|
started = time.time()
|
|
self._sample(started)
|
|
# Keep the cadence steady regardless of how long the reads took.
|
|
self._stop.wait(max(0.0, self.interval - (time.time() - started)))
|
|
|
|
def _sample(self, now):
|
|
with self._state_lock:
|
|
rcu_powered = self._rcu_powered
|
|
|
|
# With the relay open the RCU is dead: every read would burn a full
|
|
# serial timeout and stall the sampling cadence, so skip it entirely
|
|
# and log the row with the RCU columns blank.
|
|
rcu = self.rcu.read_data() if rcu_powered else dict(BLANK_RCU_DATA)
|
|
tpu = self.tpu.read_data()
|
|
relay = self.relay.read_data()
|
|
|
|
criteria_met = (
|
|
rcu['main_state'] == RCU_STATE_CHARGING
|
|
and rcu['output_current'] is not None
|
|
and rcu['output_current'] > self.current_threshold
|
|
)
|
|
|
|
with self._state_lock:
|
|
if criteria_met:
|
|
if self._ok_since is None:
|
|
self._ok_since = now
|
|
else:
|
|
self._ok_since = None
|
|
dwell = 0.0 if self._ok_since is None else now - self._ok_since
|
|
|
|
cycle, phase, note = self._cycle, self._phase, self._note
|
|
jig_x, jig_y, jig_z = self._jig
|
|
relay_cmd, charging_cmd = self._relay_cmd, self._charging_cmd
|
|
powered = self._rcu_powered
|
|
self._latest = {'rcu': rcu, 'tpu': tpu, 'relay': relay}
|
|
|
|
self.recorder.write([
|
|
datetime.datetime.now().strftime("%Y-%m-%d %H:%M:%S.%f")[:-3],
|
|
round(now - self.t0, 1), cycle, phase,
|
|
jig_x, jig_y, jig_z, relay_cmd, charging_cmd, int(powered),
|
|
rcu['main_pre_state'], rcu['main_state'],
|
|
rcu['output_current'], rcu['output_voltage'], rcu['dc_bus_voltage'],
|
|
rcu['error_code'], rcu['shadow_error'],
|
|
tpu['main_pre_state'], tpu['main_state'],
|
|
tpu['output_current'], tpu['output_voltage'], tpu['dc_bus_voltage'],
|
|
tpu['error_code'], tpu['shadow_error'], tpu['rcu_error_code'],
|
|
relay['main_state'], round(dwell, 1), note,
|
|
])
|
|
|
|
logger.info(
|
|
" [%5.1fs] RCU state=%s I=%s A V=%s V | TPU state=%s I=%s A "
|
|
"| relay state=%s | dwell=%.0fs",
|
|
now - self.t0, rcu['main_state'], rcu['output_current'],
|
|
rcu['output_voltage'], tpu['main_state'], tpu['output_current'],
|
|
relay['main_state'], dwell)
|
|
|
|
|
|
# ---------------------------------------------------------------------------
|
|
# Test sequence
|
|
# ---------------------------------------------------------------------------
|
|
def _wait(seconds, stop_when=None, poll=0.2):
|
|
"""Sleep while the monitor keeps logging. Returns True if stop_when fired."""
|
|
deadline = time.time() + seconds
|
|
while time.time() < deadline:
|
|
if stop_when is not None and stop_when():
|
|
return True
|
|
time.sleep(poll)
|
|
return False
|
|
|
|
|
|
def run_cycle(cycle, rcu, relay, jig, monitor, cfg):
|
|
"""Run one OVP cycle. Returns 'PASS' or 'TIMEOUT'."""
|
|
logger.info("=" * 62)
|
|
logger.info("CYCLE %d/%d", cycle, cfg['cycles'])
|
|
|
|
# Step 1: close the relay and let the RCU boot. The previous cycle left it
|
|
# open, so until this lands there is nothing on the other end to talk to.
|
|
# rcu_powered is only raised once the settle is over: the monitor would
|
|
# otherwise burn a full serial timeout per register on a booting RCU.
|
|
monitor.set_context(cycle=cycle, phase="relay_on", relay_cmd=1, note="")
|
|
power_up_rcu(relay)
|
|
monitor.set_context(rcu_powered=True)
|
|
|
|
# Step 2: move to the charging position and let it settle. Without a jig
|
|
# the pads are already fixed at the charging distance, so only wait.
|
|
if jig is not None:
|
|
monitor.set_context(phase="jig_near", jig=NEAR_POSITION)
|
|
jig.move_to(*NEAR_POSITION)
|
|
else:
|
|
monitor.set_context(phase="settle")
|
|
_wait(JIG_SETTLE_S)
|
|
|
|
# Step 3: start charging and wait for the criteria.
|
|
monitor.reset_dwell()
|
|
monitor.set_context(phase="charging", charging_cmd=1)
|
|
rcu.set_charging(True)
|
|
|
|
logger.info(
|
|
"Waiting for RCU main_state=%d and output current > %.1f A held for "
|
|
"%.0fs (timeout %.0fs)...",
|
|
RCU_STATE_CHARGING, cfg['current_threshold'], cfg['dwell'],
|
|
cfg['timeout'])
|
|
|
|
met = _wait(cfg['timeout'],
|
|
stop_when=lambda: monitor.dwell_seconds() >= cfg['dwell'])
|
|
result = "PASS" if met else "TIMEOUT"
|
|
|
|
if met:
|
|
logger.info("Cycle %d: criteria met after %.1fs of dwell.",
|
|
cycle, monitor.dwell_seconds())
|
|
else:
|
|
latest = monitor.latest().get('rcu', {})
|
|
logger.warning(
|
|
"Cycle %d: TIMEOUT after %.0fs (last RCU state=%s, current=%s A).",
|
|
cycle, cfg['timeout'], latest.get('main_state'),
|
|
latest.get('output_current'))
|
|
|
|
monitor.set_context(note=result)
|
|
|
|
# Step 4: open the relay. This also cuts the RCU's power, so from here on
|
|
# the RCU is dead and its telemetry columns stay blank.
|
|
relay.set_control(False)
|
|
monitor.set_context(phase="relay_off", relay_cmd=0, rcu_powered=False)
|
|
|
|
# Step 5: move away, keep monitoring. Without a jig there is nothing to
|
|
# move, so this becomes a plain observation window after the relay opens.
|
|
if jig is not None:
|
|
monitor.set_context(phase="jig_far", jig=FAR_POSITION)
|
|
jig.move_to(*FAR_POSITION)
|
|
else:
|
|
monitor.set_context(phase="observe")
|
|
_wait(FAR_OBSERVE_S)
|
|
|
|
# Step 6: stop charging and settle before the next cycle. The relay is
|
|
# already open, so this write is best-effort housekeeping - an unpowered
|
|
# RCU cannot acknowledge it, and that is not a test failure.
|
|
try:
|
|
rcu.set_charging(False)
|
|
except Exception as exc:
|
|
logger.info("Charging cmd = 0 not acknowledged, as expected with the "
|
|
"relay open and the RCU unpowered (%s).", exc)
|
|
monitor.set_context(phase="charging_off", charging_cmd=0)
|
|
_wait(CYCLE_GAP_S)
|
|
monitor.set_context(note="")
|
|
|
|
logger.info("Cycle %d finished: %s", cycle, result)
|
|
return result
|
|
|
|
|
|
def ovp_test(rcu_port, tpu_port, relay_port, jig_port, cycles, cfg, stamp):
|
|
os.makedirs(RESULTS_DIR, exist_ok=True)
|
|
csv_path = os.path.join(RESULTS_DIR, "ovp_test_%s.csv" % stamp)
|
|
|
|
rcu = tpu = relay = jig = None
|
|
monitor = None
|
|
recorder = None
|
|
results = []
|
|
|
|
try:
|
|
rcu = RCU(rcu_port)
|
|
tpu = TPU(tpu_port)
|
|
relay = Relay(relay_port)
|
|
jig = Jig(jig_port) if jig_port else None
|
|
if jig is None:
|
|
logger.info("No jig port given - running without the jig; keep "
|
|
"the pads fixed at the charging distance.")
|
|
|
|
# The RCU is powered through the relay, so close it and let the RCU
|
|
# boot before anything - the pre-flight probe included - talks to it.
|
|
power_up_rcu(relay)
|
|
|
|
# Pre-flight: prove every port talks before committing to a long run.
|
|
if cfg['link_test']:
|
|
probe_devices = [rcu, tpu, relay]
|
|
if jig is not None:
|
|
probe_devices.append(jig)
|
|
all_ok, any_dead = link_test(probe_devices)
|
|
if any_dead:
|
|
logger.error("A device did not answer at all - aborting. "
|
|
"Check the -r/-t/-l/-j port assignments.")
|
|
return
|
|
if all_ok:
|
|
logger.info("Link test: PASS")
|
|
else:
|
|
logger.warning("Link test: PARTIAL - some registers could not "
|
|
"be read; continuing anyway.")
|
|
|
|
# Known-safe starting point before any logging begins. The relay stays
|
|
# closed: cycle 1 re-asserts it and waits out the settle again.
|
|
rcu.set_charging(False)
|
|
|
|
recorder = CSVRecorder(csv_path)
|
|
logger.info("Logging telemetry to: %s", os.path.abspath(csv_path))
|
|
|
|
monitor = Monitor(rcu, tpu, relay, recorder, t0=time.time(),
|
|
interval=cfg['interval'],
|
|
current_threshold=cfg['current_threshold'])
|
|
monitor.set_context(cycle=0, phase="startup",
|
|
jig=(0, 0, 0) if jig is not None else None,
|
|
relay_cmd=1, charging_cmd=0, rcu_powered=True)
|
|
monitor.start()
|
|
|
|
for cycle in range(1, cycles + 1):
|
|
results.append(
|
|
run_cycle(cycle, rcu, relay, jig, monitor, cfg))
|
|
|
|
except KeyboardInterrupt:
|
|
logger.warning("Interrupted by user.")
|
|
except minimalmodbus.ModbusException as exc:
|
|
logger.error("Modbus error: %s", exc)
|
|
except Exception as exc:
|
|
logger.exception("Unexpected error: %s", exc)
|
|
finally:
|
|
if monitor is not None:
|
|
monitor.stop()
|
|
monitor.join(timeout=cfg['interval'] + SERIAL_TIMEOUT * 4)
|
|
|
|
# Always leave the rig de-energised.
|
|
if rcu is not None:
|
|
try:
|
|
rcu.set_charging(False)
|
|
except Exception as exc:
|
|
# Expected when we stop with the relay already open.
|
|
logger.info("Charging cmd = 0 not acknowledged (RCU likely "
|
|
"unpowered): %s", exc)
|
|
if relay is not None:
|
|
try:
|
|
relay.set_control(False)
|
|
except Exception as exc:
|
|
logger.warning("Could not switch relay off: %s", exc)
|
|
|
|
for dev in (rcu, tpu, relay, jig):
|
|
if dev is not None:
|
|
dev.close()
|
|
if recorder is not None:
|
|
recorder.close()
|
|
|
|
# Nothing to summarise if we bailed out before logging started.
|
|
if recorder is not None:
|
|
passed = results.count("PASS")
|
|
logger.info("=" * 62)
|
|
logger.info("Cycles run: %d/%d | PASS: %d | TIMEOUT: %d",
|
|
len(results), cycles, passed, len(results) - passed)
|
|
for i, res in enumerate(results, start=1):
|
|
logger.info(" cycle %d: %s", i, res)
|
|
logger.info("Results saved to: %s", os.path.abspath(csv_path))
|
|
|
|
|
|
# ---------------------------------------------------------------------------
|
|
def main():
|
|
parser = argparse.ArgumentParser(
|
|
description="BE OVP cycling test across RCU, TPU, relay and jig.")
|
|
parser.add_argument("-r", "--rcu", required=True,
|
|
help="COM port for the RCU (e.g. COM3)")
|
|
parser.add_argument("-t", "--tpu", required=True,
|
|
help="COM port for the TPU")
|
|
parser.add_argument("-l", "--relay", required=True,
|
|
help="COM port for the relay board")
|
|
parser.add_argument("-j", "--jig", default=None,
|
|
help="COM port for the jig. Omit to run without one: "
|
|
"the pads must then already be fixed at the "
|
|
"charging distance, and the move-away phase "
|
|
"becomes a plain observation window.")
|
|
parser.add_argument("-n", "--cycles", type=int, default=None,
|
|
help="Number of test cycles "
|
|
"(not needed with --link-test)")
|
|
parser.add_argument("--link-test", action="store_true",
|
|
help="Read a few registers from every port, report "
|
|
"per port, then exit without running any cycles")
|
|
parser.add_argument("--no-link-test", action="store_true",
|
|
help="Skip the pre-flight link test that normally "
|
|
"runs before the first cycle")
|
|
parser.add_argument("--current-threshold", type=float,
|
|
default=CHARGING_CURRENT_THRESHOLD_A,
|
|
help="Output current the RCU must exceed, in amps "
|
|
"(default: %(default)s)")
|
|
parser.add_argument("--dwell", type=float, default=CHARGING_DWELL_S,
|
|
help="Seconds the criteria must hold "
|
|
"(default: %(default)s)")
|
|
parser.add_argument("--timeout", type=float, default=CHARGING_TIMEOUT_S,
|
|
help="Seconds before a cycle is abandoned "
|
|
"(default: %(default)s)")
|
|
parser.add_argument("--interval", type=float, default=MONITOR_INTERVAL_S,
|
|
help="Telemetry sampling interval in seconds "
|
|
"(default: %(default)s)")
|
|
args = parser.parse_args()
|
|
|
|
if not args.link_test:
|
|
if args.cycles is None:
|
|
parser.error("-n/--cycles is required unless --link-test is given")
|
|
if args.cycles < 1:
|
|
parser.error("--cycles must be at least 1")
|
|
if args.dwell > args.timeout:
|
|
parser.error("--dwell cannot exceed --timeout")
|
|
|
|
os.makedirs(RESULTS_DIR, exist_ok=True)
|
|
|
|
# One log per run, stamped like its CSV so the two pair up by name. A
|
|
# link test writes its own file with a _linktest suffix, so a quick probe
|
|
# never lands in the middle of a real run's log.
|
|
stamp = datetime.datetime.now().strftime("%Y%m%d_%H%M%S")
|
|
log_name = "ovp_test_%s%s.log" % (stamp,
|
|
"_linktest" if args.link_test else "")
|
|
log_path = os.path.join(RESULTS_DIR, log_name)
|
|
|
|
logging.basicConfig(
|
|
level=logging.INFO,
|
|
format="%(asctime)s - %(levelname)s - %(message)s",
|
|
datefmt="%Y-%m-%d %H:%M:%S",
|
|
handlers=[logging.FileHandler(log_path, encoding="utf-8"),
|
|
logging.StreamHandler(sys.stdout)])
|
|
logger.info("Logging this run to: %s", os.path.abspath(log_path))
|
|
|
|
if args.link_test:
|
|
ok = run_link_test(args.rcu, args.tpu, args.relay, args.jig)
|
|
sys.exit(0 if ok else 1)
|
|
|
|
cfg = {
|
|
'cycles': args.cycles,
|
|
'current_threshold': args.current_threshold,
|
|
'dwell': args.dwell,
|
|
'timeout': args.timeout,
|
|
'interval': args.interval,
|
|
'link_test': not args.no_link_test,
|
|
}
|
|
ovp_test(args.rcu, args.tpu, args.relay, args.jig, args.cycles, cfg,
|
|
stamp)
|
|
|
|
|
|
if __name__ == "__main__":
|
|
main()
|